Tag: loading

Incinerator with Capacity of destruction in weight: 60 Kg/h.

Ability of destruction in weight: 60 Kg/h.

It should be in a position to operate not less than 10 hours/day

This incinerator must be able to destruct all combustible wastes generated by hospitals, private clinics, labs, institutes, etc…

Design Specification: Forms A, B, C, D, and E of medical waste 

“PYROLYTIC” combustion, by simply controlling the gasification of waste.

The incinerator must prevent the discharge of smoke and fine dust, (Smokeless) throughout the loadings.

It should be able to decrease the volume of wastes from 98%.

It should be in a position to hold emission in the next burn with gas residence of not less than 2 seconds.

The incineration must be completely free of visible smoke in addition to offensive odours.

The reduced volt electricity (L.C.P) of this waste will be 3,500 kcal/kg

The Temperatures of combustion: Minimum will be 850oC and maximum 1400oC

Post combustion: 1100oC.

The Internal diameter of the Chimney: Ø 400 and its elevation: 8 m

The Loudness of the combustion room: 1.200 L

The Dimension of this door for loading in cm: 70×70.

Burner operation should be Automatic On/Off

Gas: petrol

The supplier must provide  necessary information for the top of  this installment

This incinerator with”PYROLYTIC” combustion must have:

A combustion chamber of waste:

* Totally tight door for your manual loading of waste.  The loading must be Manual, Batch Load 

* A burner of light which the usage is limited to the ignition of waste.

* Frontage of loading door seals gone on hinges, wheel of screw plug, elastic joint, and stuffing insulating from refractory.

* The insulating material of the combustion chamber should be composed of refractory bricks, so using a high content of aluminum and insulates bricks so as to assure a minimum temperature on   the exterior sheet metal. Thickness : ≥100 mm

. Nature: 42% of Al203

Insulate in fibrous panels:

. Thickness: ≥75 mm

Nature: Calcium silicate.

* Burner of light of waste, together with gas, regular mono-bloc casting directing stirring flame, lighting and safety of digital ignition, permanent venting, electromagnetic sluice gate of regulation and isolating valve.

. Thickness: ≥75 mm

Nature: Calcium silicate.

* Burner of lighting of waste, with fuel, standard mono-bloc casting guiding  plunging flame, lighting and safety of electronic ignition, permanent ventilation, electromagnetic sluice gate of regulation and isolating valve.

* Plate of combustion in Carborundum, avoiding the fixing of glass and slags. 

Incinerator with Ease of Jealousy in weight: 60 Kg/h.

Ability of destruction in weight: 60 Kg/h.

It should be able to work less 10 hours/day

This incinerator must have the ability to destruct all combustible wastes produced by hospitals, private practices, labs, institutes, etc…

Design Specification: Types A, B, C, D, and E of medical waste 

“PYROLYTIC” combustion, by simply controlling the gasification of waste.

The incinerator must prevent the discharge of smoke and dust that is fine, (Smokeless) during the loadings.

It should have the ability to reduce the quantity of wastes from 98%.

It should be able to hold emission in the next burn with gas residence of less than two seconds.

The incineration should be completely free of visible smoke as well as offensive odours.

The lower volt electricity (L.C.P) of the waste will be 3,500 kcal/kg

The Temperatures of combustionMinimum will be 850oC and maximum 1400oC

Post combustion: 1100oC.

The Inner diameter of the Chimney: Ø 400 and its elevation: 8 m

The Volume of the combustion chamber: 1.200 L

The Dimension of this doorway for loading cm: 70×70.

Burner operation should be Automatic On/Off

Gas: diesel

The provider must supply necessary information for the top of  this installment

This incinerator with”PYROLYTIC” combustion must have:

A combustion chamber of waste:

* Totally tight doorway for the manual loading of waste.  The loading should be Manual, Batch Load 

* A burner of lighting that the use is restricted to the ignition of waste.

* Frontage of loading with door seals ended up on hinges, wheel of screw plug, elastic joint, and stuffing insulating from refractory.

* The insulation of the combustion chamber should be made up of refractory bricks, so using a high content of aluminium and insulates bricks in order to guarantee a minimum temperature on   the exterior sheet metal.

* Composition of this refractory;

Refractory concrete:

. Nature: 42 percent of Al203

Insulate in fibrous panels:

. Thickness: ≥75 mm

Nature: Calcium silicate.

* Burner of lighting of waste, with fuel, standard mono-bloc casting directing stirring flame, lighting and safety of digital ignition, permanent ventilation, electromagnetic sluice gate of regulation and isolating valve.

* Burner of lighting of waste, with fuel, standard mono-bloc casting guiding  plunging flame, lighting and safety of electronic ignition, permanent ventilation, electromagnetic sluice gate of regulation and isolating valve.

* Plate of combustion in Carborundum, avoiding the fixing of glass and slags. 

Incinerator with Capacity of destruction in weight: 60 Kg/h.

Capacity of destruction in weight: 60 Kg/h.

It should be able to operate not less than 10 hours/day

This incinerator must be able to destruct all combustible wastes produced by hospitals, private clinics, laboratories, institutes, etc…

Design Specification : Types A, B, C, D, and E of medical waste 

“PYROLYTIC” combustion, by controlling the gasification of waste.

The incinerator must avoid the release of black smoke and fine dust, (Smokeless) during the loadings.

It should be able to reduce the volume of wastes by 98%.

It should be able to hold emission in the second burn with gas residence of not less than 2 seconds.

The incineration should be completely free from visible smoke as well as offensive odours.

The lower calorific power (L.C.P) of this waste will be 3,500 kcal/kg

The Temperatures of combustion: Minimum will be 850oC and max 1400oC

Post combustion: >1100oC.

The Internal diameter of the Chimney: Ø 400 and its height: 8 m

The Volume of the combustion chamber: 1.200 L

The Dimension of the door for loading in cm: 70×70.

Burner operation should be Automatic On/Off

Fuel : diesel

The supplier must provide  necessary information for the best of  the installation

This incinerator with “PYROLYTIC” combustion must have:

A combustion chamber of waste:

* Perfectly tight door for the manual loading of waste. The loading should be Manual, Batch Load 

* A burner of lighting which the use is limited to the ignition of waste.

* Frontage of loading with door seals gone up on hinges, wheel of screw plug, flexible joint, and stuffing insulating out of refractory.

* The insulation of the combustion chamber should be composed of refractory bricks, having a high content of aluminium and insulates bricks in order to assure a minimum temperature on   the outside sheet metal.

* Composition of the refractory;

Refractory concrete :

. Thickness : ≥100 mm

. Nature: 42% of Al203

Insulate in fibrous panels :

. Thickness: ≥75 mm

Nature: Calcium silicate.

* Burner of lighting of waste, with fuel, standard mono-bloc casting guiding  plunging flame, lighting and safety of electronic ignition, permanent ventilation, electromagnetic sluice gate of regulation and isolating valve.

* Plate of combustion in Carborundum, avoiding the fixing of glass and slags. 

Incinerator with Capacity of destruction in weight: 60 Kg/h.

Capacity of destruction in weight: 60 Kg/h.

It should be able to operate not less than 10 hours/day

This incinerator must be able to destruct all combustible wastes produced by hospitals, private clinics, laboratories, institutes, etc…

Design Specification : Types A, B, C, D, and E of medical waste 

“PYROLYTIC” combustion, by controlling the gasification of waste.

The incinerator must avoid the release of black smoke and fine dust, (Smokeless) during the loadings.

It should be able to reduce the volume of wastes by 98%.

It should be able to hold emission in the second burn with gas residence of not less than 2 seconds.

The incineration should be completely free from visible smoke as well as offensive odours.

The lower calorific power (L.C.P) of this waste will be 3,500 kcal/kg

The Temperatures of combustion: Minimum will be 850oC and max 1400oC

Post combustion: >1100oC.

The Internal diameter of the Chimney: Ø 400 and its height: 8 m

The Volume of the combustion chamber: 1.200 L

The Dimension of the door for loading in cm: 70×70.

Burner operation should be Automatic On/Off

Fuel : diesel

The supplier must provide  necessary information for the best of  the installation

This incinerator with “PYROLYTIC” combustion must have:

A combustion chamber of waste:

* Perfectly tight door for the manual loading of waste. The loading should be Manual, Batch Load 

* A burner of lighting which the use is limited to the ignition of waste.

* Frontage of loading with door seals gone up on hinges, wheel of screw plug, flexible joint, and stuffing insulating out of refractory.

* The insulation of the combustion chamber should be composed of refractory bricks, having a high content of aluminium and insulates bricks in order to assure a minimum temperature on   the outside sheet metal.

* Composition of the refractory;

Refractory concrete :

. Thickness : ≥100 mm

. Nature: 42% of Al203

Insulate in fibrous panels :

. Thickness: ≥75 mm

Nature: Calcium silicate.

* Burner of lighting of waste, with fuel, standard mono-bloc casting guiding  plunging flame, lighting and safety of electronic ignition, permanent ventilation, electromagnetic sluice gate of regulation and isolating valve.

* Plate of combustion in Carborundum, avoiding the fixing of glass and slags. 

Supply, Installation and Commissioning of DIESEL FIRED INCINERATORS

Supply, Installation and Commissioning of DIESEL FIRED INCINERATORS
Medical Waste Incinerator, 100 to 120 Kg/hr
Application   For incineration, general and pathological
Capacity    100 C 120 kg/h burn rate
Type Two  combustion chambers type; primary  and Secondary, controlled/forced combustion air type with a flue gas emission scrubbing unit
Operating time                Minimum 8 hours daily
Operating temperature     From 850 0C to 1200 0C, Automatic controlled
Residual Ash                    5 to 10%
Construction Constructed from heavy duty mild or aluminized  steel
Or equal and approved equivalent

Insulation material            Refractory material lining similar or equal to calcium 
Silicate and hot face combination of heavy duty brickwork
Internal Construction        Fixed hearth type complete with gratings, concave bottom and charging door, lined with refractory material
Charging Door                   Suitable for manual loading of wastes and with smooth 
Dear seal equivalent of Ceramic seals with hinges.
Door Lock                          Automatic, Electric type
Ash removal door    Provided, for removing resultant bottom ash leftovers                              from the Primary chamber
Gratings    Provided
Loading Manual loading of waste
Primary Burner                        Fully automatic, with fuel, temperature and speed  controls with ignition system  flame detector                                                         Air fan Complete with safety features, flame failure                                                                    Diesel fired fuel injector type                                               
Flange mounted
Blower   Provided.  3 phase for supplying excess combustion air through the distribution system with speed control system
Temperature Minimum exit 850 0C
Observation port To be provided with protective glass type
     3.3    Secondary chamber

Supply, Installation and Commissioning of DIESEL FIRED INCINERATORS

Supply, Installation and Commissioning of DIESEL FIRED INCINERATORS
Medical Waste Incinerator, 100 to 120 Kg/hr
Application   For incineration, general and pathological
Capacity    100 C 120 kg/h burn rate
Type Two  combustion chambers type; primary  and Secondary, controlled/forced combustion air type with a flue gas emission scrubbing unit
Operating time                Minimum 8 hours daily
Operating temperature     From 850 0C to 1200 0C, Automatic controlled
Residual Ash                    5 to 10%
Construction Constructed from heavy duty mild or aluminized  steel
Or equal and approved equivalent

Insulation material            Refractory material lining similar or equal to calcium 
Silicate and hot face combination of heavy duty brickwork
Internal Construction        Fixed hearth type complete with gratings, concave bottom and charging door, lined with refractory material
Charging Door                   Suitable for manual loading of wastes and with smooth 
Dear seal equivalent of Ceramic seals with hinges.
Door Lock                          Automatic, Electric type
Ash removal door    Provided, for removing resultant bottom ash leftovers                              from the Primary chamber
Gratings    Provided
Loading Manual loading of waste
Primary Burner                        Fully automatic, with fuel, temperature and speed  controls with ignition system  flame detector                                                         Air fan Complete with safety features, flame failure                                                                    Diesel fired fuel injector type                                               
Flange mounted
Blower   Provided.  3 phase for supplying excess combustion air through the distribution system with speed control system
Temperature Minimum exit 850 0C
Observation port To be provided with protective glass type
     3.3    Secondary chamber

hot medical waste disposing machine, Operation Condition: 8-16 hr/ day

hot medical waste disposing machine, Operation Condition: 8-16 hr/ day 
Control: Built-in data recording
Operating temperature: Incinerator 
/Primary Combustion Chamber
Primary Chamber: 900 – 1200 0C

Type: continuous loading, Top Loading
Capacity/Burn rate per hour: 250 -300 kg/hr 
Material:
External- 3 layers 
Internal lining:  a fire proof material of pre-fired refractory bricks with Aluminium lining, resistant to corrosive waste or gas and to thermal shock
Secondary 
Combustion 
Chamber:

Type: horizontal/vertical
Temperature:  1200- 1300 oC
Residence time of gases : >2 seconds
Ash Residue: <5% of original waste size
Ash Handling System: Both Automatic and manual removal of Ash. Must ensure removal/treatment of hazardous remnants of ash
Flue gas treatment   system : Capable of treating the flow of flue  gas as the incinerator is operating at its maximum capacity
Auxiliary device: Water level gauge, pressure sensor, PH sensor..etc 
Auxiliary device: Fuel cut-off device
Waste  feeding mechanism: Automatic pneumatic/hydraulic waste loading system or conveyor belt , capacity 650-800 L at a time                                                                                                                                      
Chimney (Stack):
Type: Vertical type
height:>7 meter’
Material: Fireproof cast, stainless steel
OUTPUT: 
GAS- SMOKELESS,ODORLESS 
ASH -Max <5% of original waste size
Reduction of Pollutant gas SO2, HCL, HF and line particulate
Emission standard:
WHO/ European
Test report for emission testing provided?
Heat exchange mode: Automatic

Accessories:  
All standard accessories for incinerator, including but not limited to loading system, heat exchangers, pollution control system, ash removal system, including ladder and oil tanker (2500litre capacity).
Operating Environment: 
The incinerator is capable to operate at the altitude of 2400mt above sea level. (according to the site conditions)
Power Requirement:
220 Vac single phase or 380 Vac three phase 4 wire system  50HZ   
Installation Testing and Commissioning :   to be conducted by certified or qualified personnel.  
Supplier shall provide the following documentation    
User (Operating) manual in English. 
Service (Technical / Maintenance) manual in English. 
Certificate of calibration and inspection from factory. 
Fast moving spare parts: Supplier is able to provide fast moving spare parts with quantities as described in the price schedule or their equivalent. 
Training: Supplier is able to provide training on operation, management and maintenance of incinerators.   
Warranty 
Comprehensive warranty for minimum 2 year. 
Maintenance Service during Warranty Period 
During warranty period supplier must ensure, corrective/breakdown maintenance whenever required. 
Supplier has a local agent or branch office in Ethiopia (please indicate the name and contact details of the agent) 
Supplier provides labor, lifting equipment and tools needed for the installation, testing and commissioning of the incinerator. The supplier will be responsible for the safety of its staff, workers and laborers during the installation, testing & commisssioning 
Able to provides all necessary information that would be used as an input for preparation of floor (platform) and room for the incinerator to be supplied such as the following; 
Lay outs (drawings) and  pictures of the incinerators to be installed 
 Length, width and height of incinerator,
 Area (length x width) for the floor (platform) and Length, width and height for roofing, for each type of incinerators to be installed 
Needs during the transport, installation, assembly, commissioning and operation of the equipment in terms of access points, available space to operate (inside and outside of the room), ways to get to the site where the equipment will be placed or any others. ; 
Technical Specifications (fuel consumption rate, weight, anchoring system, support points, etc.);
 Chimney specifications to be used for the roof design ( size, support needs, insulation); 
Considerations related to operation and Maintenance, for e.g., minimum available area to perform routine maintenance tasks and replacement of key parts that need to be changed more frequently; 
Any consideration/need of the incinerator while being operated that might impact the design of the room/shelter where it will be placed.
Other important information to be considered that can influence or impact the designing and building processes of the infrastructure for the incinerators, for example area dimension for other accessary parts like fuel tanker storage, if applicable, etc…. 
Delivery time (DAP, Addis Ababa) in weeks
Name of manufacturer including website of the manufacturer &/ supplier & country of origin

hot medical waste disposing machine, Operation Condition: 8-16 hr/ day

hot medical waste disposing machine, Operation Condition: 8-16 hr/ day 
Control: Built-in data recording
Operating temperature: Incinerator 
/Primary Combustion Chamber
Primary Chamber: 900 – 1200 0C

Type: continuous loading, Top Loading
Capacity/Burn rate per hour: 250 -300 kg/hr 
Material:
External- 3 layers 
Internal lining:  a fire proof material of pre-fired refractory bricks with Aluminium lining, resistant to corrosive waste or gas and to thermal shock
Secondary 
Combustion 
Chamber:

Type: horizontal/vertical
Temperature:  1200- 1300 oC
Residence time of gases : >2 seconds
Ash Residue: <5% of original waste size
Ash Handling System: Both Automatic and manual removal of Ash. Must ensure removal/treatment of hazardous remnants of ash
Flue gas treatment   system : Capable of treating the flow of flue  gas as the incinerator is operating at its maximum capacity
Auxiliary device: Water level gauge, pressure sensor, PH sensor..etc 
Auxiliary device: Fuel cut-off device
Waste  feeding mechanism: Automatic pneumatic/hydraulic waste loading system or conveyor belt , capacity 650-800 L at a time                                                                                                                                      
Chimney (Stack):
Type: Vertical type
height:>7 meter’
Material: Fireproof cast, stainless steel
OUTPUT: 
GAS- SMOKELESS,ODORLESS 
ASH -Max <5% of original waste size
Reduction of Pollutant gas SO2, HCL, HF and line particulate
Emission standard:
WHO/ European
Test report for emission testing provided?
Heat exchange mode: Automatic

Accessories:  
All standard accessories for incinerator, including but not limited to loading system, heat exchangers, pollution control system, ash removal system, including ladder and oil tanker (2500litre capacity).
Operating Environment: 
The incinerator is capable to operate at the altitude of 2400mt above sea level. (according to the site conditions)
Power Requirement:
220 Vac single phase or 380 Vac three phase 4 wire system  50HZ   
Installation Testing and Commissioning :   to be conducted by certified or qualified personnel.  
Supplier shall provide the following documentation    
User (Operating) manual in English. 
Service (Technical / Maintenance) manual in English. 
Certificate of calibration and inspection from factory. 
Fast moving spare parts: Supplier is able to provide fast moving spare parts with quantities as described in the price schedule or their equivalent. 
Training: Supplier is able to provide training on operation, management and maintenance of incinerators.   
Warranty 
Comprehensive warranty for minimum 2 year. 
Maintenance Service during Warranty Period 
During warranty period supplier must ensure, corrective/breakdown maintenance whenever required. 
Supplier has a local agent or branch office in Ethiopia (please indicate the name and contact details of the agent) 
Supplier provides labor, lifting equipment and tools needed for the installation, testing and commissioning of the incinerator. The supplier will be responsible for the safety of its staff, workers and laborers during the installation, testing & commisssioning 
Able to provides all necessary information that would be used as an input for preparation of floor (platform) and room for the incinerator to be supplied such as the following; 
Lay outs (drawings) and  pictures of the incinerators to be installed 
 Length, width and height of incinerator,
 Area (length x width) for the floor (platform) and Length, width and height for roofing, for each type of incinerators to be installed 
Needs during the transport, installation, assembly, commissioning and operation of the equipment in terms of access points, available space to operate (inside and outside of the room), ways to get to the site where the equipment will be placed or any others. ; 
Technical Specifications (fuel consumption rate, weight, anchoring system, support points, etc.);
 Chimney specifications to be used for the roof design ( size, support needs, insulation); 
Considerations related to operation and Maintenance, for e.g., minimum available area to perform routine maintenance tasks and replacement of key parts that need to be changed more frequently; 
Any consideration/need of the incinerator while being operated that might impact the design of the room/shelter where it will be placed.
Other important information to be considered that can influence or impact the designing and building processes of the infrastructure for the incinerators, for example area dimension for other accessary parts like fuel tanker storage, if applicable, etc…. 
Delivery time (DAP, Addis Ababa) in weeks
Name of manufacturer including website of the manufacturer &/ supplier & country of origin

Top Feeding Door Solid Waste Incinerators HICLOVER

HICLOVER Top Feeding Door Waste Incinerators

▲Large Feeding Door
▲Better for Animal Waste Cremation
▲HICLOVER Brand Section

***Primary Combustion Chamber
***Secondary Combustion Chamber
***Smoke Filter Chamber
***Optional Wet Scrubber System(1 section or  3 sections)

Phone: +86-13813931455(whatsapp/wechat)
Website: http://www.hiclover.com/ 
Nanjing Clover Medical Technology Co.,Ltd.
Items Side Loading Door Top Loading Door
Cost Cheap High cost
Manufacture Tech. Easy A litter difficult
Safty safe safer
Incinerator Shape L/ I
Loading manual big door can be by Forklift
Land occupation small big
Ash door same with feeding door or alone alone
market share Most-CHINA/USA/INDIA Few- UK/USA
Dooring open type Manual Pry bar/Hydraulic pressure
Dooring Open size small big
Better for medical bag waste animal waste
open when burning safer, can feed waste, few smoke not safe, cannot feed waste, most smoke



parts of an incinerator
physical method of disposing hazardous chemical waste
portable incinerators types
proper ways of waste disposal in the community
questionnaire to ask about incinerators for collection of data
HICLOVER Solution for Fighting COVID-19, Medical Waste Incinerator

 

2020-08-07