Tag: Blog

medical waste diesel incinerator

medical waste diesel incinerator Burning Rate:47.5–50kg/hour
Minimum Operating Temperature:850o C(primary combustion)
Maximum Operating Temperature:1450 oC(Afterburner chamber)
Secondary Burning Capacity:Compulsory
Density Insulation:20mm High
Refractory Casting:Minimum 65mm
Heavy duty refracting lining to withstand:1500 oC
Average burn out time :maximum:3 hrs
Fuel type:Diesel
Fuel Consumption in L/hr:7.5 to 9
Residence time in secondary chamber:2sec
Residue ash post 100kg:Max 3.8kg
Device for preventing infectious spatter and/or cross-contamination by safely sterilizing needles, rings, etc. by infrared heat.
Electrical characteristics
1) A 120V, 60Hz mono-phase electrical source with line connection plug NEMA 5-15 type.
2) Protections against over-voltage and over-current line conditions.
3) Compliance with applicable Haitian standards and regulations.

Operational characteristics
4) Fast complete sterilization in around 5-7 s.
5) Fully isolated quartz/ceramic tube, to grant safety for the user (no open flame).
6) Working temperature around 820°C.
7) Mounted on a stable base, as per example below.

Incinerator with 1 or 3 section type wet scrubber

Medical waste is currently managed via an old medical waste incinerator; which also manages the associated medical waste delivery and ash removal arrangements.Pathological waste: organs, tissues, surgical specimens, bodily fluids removed during surgery or autopsy.  Mix and moisture content not known.Allow for weekly deliveries @ 8 hour per day operation = 11 kg per hour capacity.

Model

YD-150

Picture

Burning Rate

Average 150 kgs/hour

Feed Capacity

Average 300 kgs/feeding

Primary Combustion Chamber

1500 Liters

Secondary Combustion Chamber

500 Liters

Mix Combustion Chamber

Yes

Smoke Filter Chamber

Wet Scrubber System(3 Section Type)

Feed Mode

Manual

Voltage

380V

Power

9.0 Kw

Fuel Type

Diesel Oil/ Natural Gas/LPG

Burner

Italy Original

Oil Consumption (Diesel Oil)

Average 24 kg/hour

Gas Consumption (Natural Gas)

***

Internal Dimensions

150 x 100 x 96cm (primary chamber)

External Dimensions

240 x 170 x 380cm(without chimney/wet scrubber)

Temperature Monitor

Yes

Oil Tank Capacity(if oil fuel)

300 Liters

Door Opening

59 x 81cm

Chimney Length

10.0 Meters

Chimney Type

Stainless Steel

Equipment Gross Weight

11000 kgs

Operation Technical Specifications

Primary Chamber Temperature

800-1000

Secondary Chamber Temperature

1000-1200

Residency Time

2.0 Sec.

Burning Efficiency

98%

Waste Lower Calorific Power

3000Kcal

HICLOVER – Medical Environmental 


 

Waste Incinerators
Medical Waste Incinerator
Pet Animal Cremation
Solid Waste Incinerator

Tel:  +86-25-8461 0201   
Mobile: +86-13813931455(whatsapp/wechat)
Website: www.hiclover.com  
Email: [email protected]
Email: [email protected]  
Nanjing Clover Medical Technology Co.,Ltd.

 

2020-06-27

waste incinerator manufacturers from china

Incinerate: medical waste and general waste Capacity to burn waste:
400-500kg/hr
Fuel: Diesel
Fuel Consumption: 10 liters/hr maximum
Incineration temperature: 950-1300 degrees centigrade
Casing: thick (min 5mm) stainless steel casing and fully insulated to
withstand sea humidity Voltage 220/240v or 380/415V Frequency 50Hz
high-capacity, powerful, European Standard incinerator,ideal for high volume incineration,with integral after-burner for greater efficiency and emissions control.
Waste Type and Burn Rate
Designed specifically for use by poultry farmers,
sheep farm owners,   abbatoirs and for pet
cremation. The incinerator will burn up to 50kg of
waste per hour (up to 150kg burn
Construction
Mild steel welded fabrication consisting of sheet steel
and structural section supports.
Lining
Monolithic high grade refractory concrete with high grade insulation backing.
Burners
Fully automatic, high-efficiency burners with electronic ignition, flame recognition and combustion control devices fitted.
Loading – Ash removal
Manual ash removal from the primary chamber. (Ash doors are available on request).
Exterior finish
Multi-coat high quality, high-temperature aluminium enamel top coat paint, finished in silver.
Control
Fully automatic sequential control provided by internal
waste incinerator manufacturers from china, PLC system providing control of:
1. Secondary chamber pre-heat
2. Burn cycle (variable)
3. Cool down cycle
Fuel Options
Diesel, LPG, Natural or Bio Gas.
Typical Fuel Consumption
Diesel: 8 / 10 litres per hour
Liquid Propane Gas: 6 / 8 litres per hour
Natural Gas: 7 / 9 Nm3 per hour
Electrical supply
240volt 16 amp.
Key Specifications
Chamber Volume 2.38m3:
2.2m (l) x 0.9m(w) x 1.2m(h)
Load Capacity up to 1000kg
Load Method Top.·         Type: Pyrolytic

·         Use: For drugs

·         With Chassis to be Mounted in truck/lorry

·         May have a primary and secondary combustion chamber

·         Must have a dilution chamber

·         Capacity: 100 Kgs per hour

·         Power Source: Diesel or Natural Gas or Liquified Natural Gas

DIESEL FIRED INCINERATOR, WITH 1000 LTR FUEL TANK AND 6M CHIMNEY

Items/Model

TS250(PLC)

Picture

Principle

Waste Incineration Treatment

Burn Rate

Average 250 kg/hour

Feed Capacity

Average 400kg/feeding

Control Mode

PLC(Programmable Logic Controller Incinerator)

Combustion Chamber

2000L

Internal Dimensions

210x88x110cm

Secondary Chamber

1000L

Smoke Filter Chamber

Yes

Feed Mode

Manual

Voltage

220V

Power

2.0Kw

Oil Consumption (kg/hour)

Average 24.2(DO)

Gas Consumption (m3/hour)

***

Temperature Monitor

Digital Display

Temperature Protection

Yes

Oil Tank

1000L

Feed Door

2.04 x 1.06 M

Chimney

3Meter

Chimney Type

Stainless Steel

1st. Chamber Temperature

800-1000

2nd. Chamber Temperature

1000-1200

Residency Time

2.0 Sec.

Gross Weight

9000kg

External Dimensions

3.00mx1.60mx3.00m (without chimney)

INCINERATION IS A WASTE TREATMENT PROCESS THAT INVOLVES THE COMBUSTION OF ORGANIC SUBSTANCES CONTAINED IN WASTE MATERIALS. INCINERATION AND OTHER HIGH-TEMPERATURE WASTE TREATMENT SYSTEMS ARE DESCRIBED AS “THERMAL TREATMENT”. INCINERATION OF WASTE MATERIALS CONVERTS THE WASTE INTO ASH, FLUE GAS AND HEAT. THE ASH IS MOSTLY FORMED BY THE INORGANIC CONSTITUENTS OF THE WASTE AND MAY TAKE THE FORM OF SOLID LUMPS OR POWDER.

TWO (2) YEAR MAINTENANCE SPARES PACKAGE FOR LINE ITEM 2 X OIL FIRED BURNER

2 X CEMENT REPAIR KIT (25KG)

2 X THERMOCOUPLE

2 X FIRE ROPE (1M)

2 X HT MASTIC (PER INCINERATOR)

1 X PPE KIT


HICLOVER – Medical Environmental 


 

Waste Incinerators
Medical Waste Incinerator
Pet Animal Cremation
Solid Waste Incinerator

Tel:  +86-25-8461 0201   
Mobile: +86-13813931455(whatsapp/wechat)
Website: www.hiclover.com  
Email: [email protected]
Email: [email protected]  
Nanjing Clover Medical Technology Co.,Ltd.

 

2020-06-27

General specifications of a 100 kg/hr three combustion Chambers Incinerator

General specifications of a 100 kg/hr incinerator
No. Description Parameter value  
  model   HICLOVER YD100
1 Machine working mode Oil + Power Oil + Power
2 Voltage 220V C 3 phase or 1 phase 220V 1P
3 Oil consumption (lit/hr) Average 25-32 lit/hr/chamber Average 33Liters/hour/total incinerator 3 burners
6 Incineration rate 100 kg/hr = 400-600 kg/8hr average 100kgs per hour
7 Feed mode Manual manual
8 Primary combustion rate 1000 lit 1080 liters
9 Secondary combustion rate 240 lit 207 Liters
10 Tertiary combustion rate 416 lit 452 liters
11 Combustion chamber material High temp. resistant concrete High temp. resistant concrete
12 Internal dimensions (1st chamber) 100*100*100 cm (l*w*h) = 1000 lit 120*100*90cm(l*w*h) = 1080 lit
13 Internal dimensions (2nd  chamber) 40*40*150 cm = 240 lit Diameter 42*150cm=207 Liters
14 Internal dimensions (3rd  chamber) 40*40*260 cm = 416 lit Diameter 60*160cm=452 Liters
15 External dimensions 300*155*200 cm (l*w*h) 220*160*370 cm (l*w*h)
16 Door opening 90*90 cm 90*90 cm
17 Additional air fan 1 fan (on the door) 1 fan (on the door)
18 Fuel type Diesel oil/ Diesel oil
19 Burner Brand: Radiator or Hofmat Italy Baltur
20 Chimney length 3-4 meters 5 Meters
21 Chimney type and material H-type stainless steel, or High temp. resistant concrete 1 Meter High temp. resistant concrete +4 Meter Stainless steel
22 Smoke and smell filter Yes, two stages. Dry Scrubber 2 stages
23 Temperature monitor Yes, (for 1st and 3rd chambers) Yes, (all 3 chambers)
24 Temperature control Optional yes
25 Oil tank 500 lit needed. yes
26 Burning efficiency >98% yes
27 1st chamber temperature 300-700 oC 800-1000 oC
28 2nd chamber temperature 500-1000 oC 1000-1200 oC
29 3rd chamber temperature 800-1200 oC 1000-1300 oC
30 Smoke residency time >2 sec. >2 sec.
31 Gross weight 5-8 tons average 6.5 Ton
32 No. of Burners 3 3


HICLOVER – Medical Environmental 


 

Waste Incinerators
Medical Waste Incinerator
Pet Animal Cremation
Solid Waste Incinerator

Tel:  +86-25-8461 0201   
Mobile: +86-13813931455(whatsapp/wechat)
Website: www.hiclover.com  
Email: [email protected]
Email: [email protected]  
Nanjing Clover Medical Technology Co.,Ltd.

 

2020-06-27

incinerator for medical waste

MINIMUM REQUIREMENTS FOR ENGINE OF INCINERATION
CAPACITY OF INCINERATION 30KG / H

1. GENERAL SPECIFICATIONS:

The incinerator for medical waste must be designated for the destruction of all types of waste produced by the Center for Health (waste type hospital, clinic, medical laboratory for destructive capacity of 30kg / h) .

Pyrolytic gasification combustion control shall:

? Prevent development of black smoke and dust formation tine loading of waste;
? Provide regular and complete combustion of the waste;
? Provide controlled guaranteeing the absence of color, cloud of smoke or burning odor;
? Allow limited fuel consumption

The incinerator with pyrolytic combustion must have:

a. Combustor:

– Chamber strong steel support legs sheet;
– The lining is made of refractory bricks and a thermal insulation ensuring a minimum temperature of the sheet;
– The sole combustion is composed of a refractory lining specially designed to meet the thermal, mechanical and corrosive liquid waste requirements and molten glass;
– The dimensions of the floor are equipped laterally primary air nozzles that fuel combustion;
– The hearth is equipped with a stirring system of waste combustion is controlled switch;
– The recovery operation of the ashes will take place once a day line before starting the day’s operation.

Composition of the refractory lining:
* Insulation:
– Operating temperature:             1000 ° C max
– Nature:                    Calcium Silicate
– Thickness:                    80 mm
* Refractory:
– Operating temperature:             1650 ° C max
– Nature:                    Refractory bricks with 42 alumina
– Thickness:                    110mm

Burner combustion:

– The combustion chamber is equipped with a burner to line to ensure correct temperature for the total destruction of waste combustion;
– Automatic monobloc burner and flame plunge equipped with an electronic ignition system and permanent ventilation.
Air fan:
– A primary air and secondary air fan provides the necessary oxygen delivery
– There is a motor operated valve that controls the flow according to the combustion.
– The pressure of the fan is 500 mm CE to pierce the mass of waste during combustion.

b. Post – combustion chamber:

– Treat the gas has a minimum temperature of 1100 ° C with a minimum residence time of two seconds the afterburner is to;
– A secondary air injection is expected and provides perfect oxygen content;
– The burner ensures a minimum temperature of 1100 ° C continuous;
– The design of this burner is similar to combustion burners;
– The afterburner is equipped with access doors in order to ensure regular cleaning of dust;
– The afterburner is lined with high thermal insulation with a thickness of 150 mm

c. Cabinet, control and regulation:
– The entire installation will conform to European standards (EU)
– The electrical power is 220V 50Hz single phase.
– Regulators of temperature with digital display:
? 1 to regulate the temperature of the combustion chamber
? 1 for controlling the temperature of the afterburner
? The manual controls are provided on each item.

2. PRODUCT DIMENSIONS

Description    Dimensions
Destructive capacity    30 Kg / hour
Operation    12h/24, 6d / 7
Fuel    diesel oil
Chimney    Height: 5m, Diameter: 280mm
Volume of combustion

chamber Combustion: 0.3 m3
Post combustion: 0.2m3
Mixed combustion chamber: 0.3m3
` TEMPERATURES:
. Combustion
. Afterburner
900 ° C
1100 ° C
POWER BURNER
Combustion
Afterburner

0.11 KW
0.11 KW0

Fan power    0.37 kW

Incinerator Parts Mineral Insulated Thermo couple sensor K-Type 300SS

Incinerator Parts Mineral Insulated Thermo couple sensor K-Type 300SS
K type 300SS: M27*2, 300mm Ins,2520 S.S, 2.5mm core 

HICLOVER – Medical Environmental 


 

Waste Incinerators
Medical Waste Incinerator
Pet Animal Cremation
Solid Waste Incinerator

Tel:  +86-25-8461 0201   
Mobile: +86-13813931455(whatsapp/wechat)
Website: www.hiclover.com  
Email: [email protected]
Email: [email protected]  
Nanjing Clover Medical Technology Co.,Ltd.

 

2020-06-27

small scale medical waste incinerator

Toxic and dangerous wastes,
Medical pharmaceutical and hospitable liquids  Wastes
Domestic garbage
Electric and electronic wastes
Industrial wastes
In General, all type of wastes and garbages
Capacity 50-60Kg/h

Combustion supporting materials
Coal oil, diesel oil

Temperature inside
800-10000C

1000-12000C

Burning efficiency
>98%

Power consumption
1.3KW

Noise
<78dB

Index of let smoke and waste gas
Ringelman’s smoke index< level 1

Smoke concentration
Oil
45.5

(mg/m3)
Gas
25.5

(mg/m3)

SO
Oil
0.176
Gas
0.065

NOX
Oil
0.031
Gas
0.038

H2S
Oil
0.016
Gas
0.008

CO
Oil
8.32
Gas
35.64

small scale medical waste incinerator Requirements
Incombustible and explosive materials not allowed in

Process manner
Two combustion cabinets (wetness drying, high temperature dust removal)

Combustion / Burner
Good Quality made in Italy

Blower
Good Quality made in China

Control Electrical
Good Quality made in China

Capacity 50 Kg/h

Working Voltage
220 V

Power
1000 W

Applied For
200 beds

Input Fashion
Manual

Medical Incinerator, 24 kg/hr
Medical Incinerator, 24 kg/hr including panel board, controls and accessories
Furnace gate
Chimney
Blower
Combustion burner

Medical Incinerator, 12 kg/hr
Medical Incinerator, 12 kg/hr including panel board, controls and accessories
Furnace gate
Chimney
Blower
Combustion burner

Medical Incinerator, 6 kg/hr
Medical Incinerator, 6 kg/hr including panel board, controls and accessories
Furnace gate
Chimney
Blower
Combustion burner

50kgs/hr iraq general waste incinerator with 2 combustion chambers

Items/Model

TS50M

Picture

 

Burn Rate

Average 50 kg/hour

Feed Capacity

Average 50-150 per feeding

Control Mode

PLC(Programmable Logic Controller Incinerator)

Combustion Chamber

450L

Internal Dimensions

75x75x80cm

Secondary Chamber

225L

Smoke Filter Chamber

Yes

Feed Mode

Manual Feeding

Voltage

220V

Power

0.5Kw

Oil Consumption (kg/hour)

Average 13.3

Gas Consumption (N.G, m3n/Hour)

Average 16

Gas Consumption (L.P.G, m3n/Hour)

Average 11.7

Temperature Monitor

Four Digital Display

Temperature Protection

Yes

Oil Tank

100L

Feed Door

55x50cm

Chimney

8Meter

Chimney Type

Stainless Steel

1st. Chamber Temperature

800-1000

2nd. Chamber Temperature

1000-1200

Residency Time

2.0 Sec.

Gross Weight

4500kg

External Dimensions

175x120x140cm

medical waste incinerator for sale

Waste type: health care waste
Waste density: 125 to 350Kg/M3
Waste calorific value: 4000KCAL/kg
Moisture content: 10 – 25%
Estimated ash residue: 5 – 10%
Nominal capacity: 50kg/hr batch load
Secondary chamber: 2 [email protected]?C
Environmental: NEMA ACT (EMCA) 1999, Waste Management regulations, 2006
Primary Combustion chamber:
The primary combustion chamber is the main structure of the incinerator, fabricated from mild steel plate and stiffened with rolled steel sections.

It is carried from the floor level on a substantial steel support frame.  Internally the shell is lined with calcium silicate insulation and a hot face combination of super duty brickwork or high strength castable.

The primary combustion chamber must ensure minimum exit temp is not less than 850?C.
One Manual Load Incinerator comprising:
Manual loading door
Horizontal solid hearth, refractory lined primary combustion chamber.  Approx. dimensions 800x2100mm(DxL)
Full sized bulk load/ash removal door
1 off diesel-fired ignition burner – temperature controlled
1 off combustion air supply fan and distribution system, automatic control
High capacity secondary combustion chamber with residence time not less than 2 seconds @>850?C
1 off diesel-fired afterburner – temperature controlled
Control panel – automatic operation
Temperature indication primary and secondary chambers 0-1200?C
Refractory lined flanged chimney to a height of 10M above ground level
NB: The secondary combustion chamber must ensure that the gas temp as measured against the inside wall in the secondary chamber & not in the flame zone, is not less than 1100?C

Throughput: 50kg/hr at a nominal calorific value of 4,000kca/kg and an average bulk density of 120kg/m3 for general waste and 160kg/m3 for health care waste.
Ash: The waste will be reduced by 90~95% to leave 10-5% ash
Quality/Environmental Standards: NEMA ACT (EMCA) 1999,   BS3316
Operating period: anticipated daily operating period 8 hours
Maximum operating period: 12 hours
Chamber: should be refractory lined to the following standard;
Hearth and lower section of chamber walls
Hot face material:
Super duty firebrick
Limiting continuous service limit 1500?C
Thickness: 100mm
Thermal conductivity: 1.3W/mk
Insulation:
Calcium silicate
Limiting continuous service limit 1000?C
Thickness: 25mm
Thermal conductivity: 0.095W/mk
Chimney/Stack: Should be manufactured from refractory lined mild steel or high quality air-cooled s/s.
The chimney/stack should have a minimum height of 10m above ground level
Ignition burners: fully packaged oil type, complete with electric ignition and flame failure controls, wired with continuous running fan and complete with all valves, fan and motor.
Maximum rating: at least 450kW/hr
Burner: arranged for on/off operation
Maximum consumption: at least 6Ltrs/hr
Average consumption: at least 3Ltrs/hr
Afterburner:
fully packaged oil type, complete with electric ignition and flame failure controls, wired with continuous running fan and complete with all valves, fan and motor.
Maximum rating: at least 450kW/hr
Burner: arranged for on/off operation
Maximum consumption: at least 9Ltrs/hr
Average consumption: at least 4.5Ltrs/hr
Flue gas treatment:  should have particulate remover (scrubbers)
Electrical supply: single phase: 220/240VAC; 50/60Hz

Documentation:
medical waste incinerator for sale Operation manual: 2 sets
Service/maintenance manual: 2 sets
Certification: NEMA (or other recognized) certification
Capacity to service equipment: the agent/supplier shall have adequate facilities, spare parts, consumables and qualified skilled technical staff.
Training: the Agent shall conduct a two days incinerators training after installation and commissioning of the equipment.

The Agent shall conduct a two days operation & maintenance training to the Biomedical engineering personnel after installation & commissioning of the equipment.